Plots from CPR tube tests : B field tests
KA2626 : -550 V  
N.D. 1.0 : roughly 1000 photoelectrons 
Definition of the
coordinate system.
Same as 
Hamamatsu 
in y and z, the opposite direction in x. 
Note the respose plots below comes with the back view
(cross talks are the front view, ok).
Will fix this later.
B field along x or y axis. 
Created by a pair of magnets. 
Field is 30- 40 Gauss, fairly uniform, 
and almost zero ( < 0.5 G ) 
for other components.
B field along z axis.
Created by magnet at the bottom of the tube case. 
Field at the outside of the cathode surface 
of the tube is about 2 Gauss (center)
and 5 Gauss (outer channels). 
Must be stronger insdie closer to the bottom, 
namely at the dynodes and anodes. 
Also x and y components may be non-negligible (up to 2-3 Gauss)
Conclusions: 
Compatible with Hamamatsu results (which is for 64-channel version, though).
Most affected by B field in negative y direction. 
Not really sensitive to B field in x (+ or -) direction. 
Can be 
sensitive in -z direction. 
Cross talk changes are totally unnoticeable 
even at 50 Gauss in x or y, and a few G in z.
Overall, 1-Gauss fields expected inside the mu-metal shields
should  not affect anything at all.
    
    Fumihiko Ukegawa